OMA Protocol Review: OMAMAN-56 Sulfites in Shrimp
(b) Results.- The results are presented in Table 1. Linearity and residual plots for the different calibration ranges are presented in Figures 2, 3, 4, and 5. The results obtained show that the method is linear. The residual plots appear to have a sigmoidal pattern or concave if the zero point is disregarded. In either case, the patterns look non-random, but the residuals are all <10%, varying from 0.2% to 3.9% for the 30-600 ppm range and 0.4% to 9.3% for the 10-300 ppm range. Thus, the response is very close to linear.
Table 1. Linearity in 30-150 ppm and 50-300 ppm ranges. 30-150 ppm
50-300 ppm
Sodium sulfite Std., ppm
Biofish result, ppm
Sodium sulfite Std., ppm
Biofish result, ppm
Mean SD Residual
Mean SD Residual
0 0 0 0 7 8 8 8
0 0 0 0
0
0
0
0.035
0
0
0
0.464
23 25 23 23 47 45 47 43 99 98 97 95
10
8
0.5
0.747
30
24 1.0 0.942
26 28 27 30 47 50 46 47 95
30
28 1.7
0.424
50
46 1.9 0.384
50
48 1.7
0.198
100
97 1.7 -0.562
142 144 147 138 294 285 278 283 525 533 521 559
102
100
98 3.1 -0.610
150
143 3.8 -0.643
97 96
148 153 147 147 285 299 275 280
150
149 2.9 -1.903
300
285 6.7 -1.680
300
285 10.3 1.108
600
535 17.1 1.095