Total Collaborative Study Protocol_Solus One Salmonella v1 1
1 2 3 4 5 6 7 8 9
automatically using an onboard barcode scanner.
1.10 If using an onboard barcode scanner slide the sample tube racks onto the machine and
click next.
1.11 Click on the Solus One Salmonella assay. Then click the “Add Assay” button.
1.12 Select “Done” when assay selection is completed. 1.13 Select the “Accept” button to start the run.
1.14 After the timeline has been accepted, loading of sample tube racks, reagents and
consumables is initiated.
1.15 If not already onboard, load sample test tube racks into the slots. 1.16 Remove the lid from the pipette boxes and load into the instrument. 1.17 Load the Conjugate, TMB Substrate, and Stop Solution into the correct locations
10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38 39 40 41
specified by the instrument.
1.18 Load the Negative and Positive Controls in each position specified by the instrument. 1.19 Prepare and dispense the correct amount of Activated Washing Buffer.
1.20 Verify all disposable containers do not need emptying.
1.21 When the run is completed, open the lid to the Dynex DS2 instrument and remove the loading trays, with the sample test tubes, by pulling the loading trays out away from the instrument. Discard sample test tubes in the proper location according to laboratory
procedures.
1.22 Discard the reagent tubes, negative controls, and positive controls in the proper location according to laboratory procedures. Note: Reagents may be capped and refrigerated at
2–8°C with the rest of the remaining kit.
1.23 To view a report on a completed run, select the desired result file in the “Recent Test”
Box, and select the “Report” button.
1.24 The report can be viewed and printed from the report display screen. 1.25 The results are displayed in a 96 well format with each individual optical density
reading.
1.26 A secondary file which can be used to export the results into a LIMS system is created in a folder on the desktop called “ELISA Export Results”, this is in a CSV (comma separated
values) file format.
1.27 Interpretation: The presence or absence of the pathogen is expressed as optical density (OD 450 ) measurements using the Dynex DS2 instrument plate reader or a microplate reader with a 450 nm filter where a Dynex DS2 instrument is unavailable. For control
assay acceptance criteria, a result of an OD 450
< 0.100 is valid for a Negative Control,
> 0.500 is valid for a Positive Control. A result of an OD 450 <
whereas a result of an OD 450
0.200 is negative for Salmonella . A result of an OD 450
≥ 0.200 is considered presumptive
positive for Salmonella .
1.28 Manual Procedure : Remove the kit from storage at 2–8°C an hour before use to allow the components to reach room temperature (15–25°C). Determine the number of wells required for the test. Take the required number of strips from the pouch and fit them to the framer provided. Note: Unused strips should be returned to the pouch and stored at
42
2–8°C
19