ESTRO 2021 Abstract Book

S582

ESTRO 2021

(P MC /P DC /P AC /P ADC /P AAC , see Fig. 1). Dose differences were quantified using dose-volume histogram (DVH)

parameters: Heart D mean

, Heart D max

, Lungs-GTV D mean

, Lungs-GTV V 5

, Lung contra

V 5

, SpinalCord 03

D max

, MedEnv 05

D max . Inter-plan variations between treatment plans generated using clinically acceptable contours were quantified dosimetrically, and the influence of contour variation on treatment plan generation and evaluation was analyzed. , and Esophagus D max

Results Clinically acceptable treatment plans could be generated fully automated for almost all patients. For only 10/140 (7%) of the DVH parameters distributed over 6 patients, the DVH parameter was above the clinical constraint (3x Heart D mean , 1x Lungs-GTV D mean , 5x Lungs-GTV V 5 , and 1x SpinalCord 03 D max ). Inter-plan variability on clinically acceptable contours was highest for Heart D max (3.4±6.8Gy) and lowest for Lungs-GTV D mean (0.3±0.4Gy). Dose differences in plan evaluation between treatment plans (P DC /P AC ) on automatic contours (C DC /C AC ), and evaluated on the three clinical contours (C MC /C ADC /C AAC ) were highest for Heart D mean (C DC : 1.8Gy, C AC : 0.6Gy), Heart D max (C DC : 6.0Gy, C AC : 2.7Gy) and Esophagus D max (C DC : 8.7Gy, C AC : 2.3Gy). Average dose differences in plan generation using the various contouring methods, evaluated on C MC , were on average below 1Gy/1%, with 11% and 15% above 2Gy/2% for P DC and P AC respectively. See Fig. 2 for an example patient with corresponding DVH’s.

Conclusion Dose differences arising from treatment plans generated on automatic contour variations were of the same (or lower) magnitude as inter-plan variability. For the heart, it is recommended to always correct delineation errors because they result in larger differences in DVH values. For the lungs, mediastinum, and esophagus,

Made with FlippingBook Learn more on our blog