SPDS 09052014
SMPR Key Points: Identification Method
• Intended Use : Reference method for dispute resolution. • Applicability : Identification of PDE5 inhibitors in dietary ingredients and supplements. • Identification Method: A method that is capable of identifying the PDE5 inhibitors listed in Annex II and providing at least class identification of other PDE5 inhibitors based on their structural similarity to the d li t d i A II Th id tifi ti h ld compoun s s e n nnex . e en ca on s ou be done using technique‐specific, generally acceptable criteria, such as those given in 2002/657/EC. • Maximum Time‐To‐Result: No maximum time.
SMPR Key Points: Identification Method
Minimum Acceptable Results 90% POI § of the pooled data for all target compounds and matrices
Target Test Concentration
Parameter
Parameter Requirements
Minimum of 33 replicates
representing all target compounds in Annex II and all matrix types listed in Annex III, spiked at or below the designated low level target test conc. Minimum of 5 replicates per matrix type spiked at 10x the
100 ppm
POI @ low concentration
POI @ high
10 x low conc .
concentration designated low level target test conc.
100% correct analyses are expected
boratory Validation
Single La
POI @ zero concentration
Minimum of 5 replicates per matrix type.
0 ppm
§ 95% confidence interval