No confirmation is required, however, if necessary colonies may be isolated for further 1
identification. Lift the top film and pick the colony from the gel. Test using appropriate 2
3
reference method procedures.
4
5
Validation Study
6
This validation study was conducted under the AOAC Research Institute Performance Tested 7
Method SM (PTM) program and the AOAC INTERNATIONAL Methods Committee Guidelines for 8
Validation of Microbiological Methods for Food and Environmental Surfaces (5). The 9
inclusivity/exclusivity study, robustness testing and matrix studies for all claimed matrixes were 10
conducted independently by Q Laboratories, Inc. (Cincinnati, OH). Product consistency and 11
12
stability data will be submitted when available.
13
14
Inclusivity/Exclusivity Testing
15
16
The objective of this experiment was to determine the ability of the 3M Petrifilm Rapid E.
coli /Coliform Count Plate method to detect a variety of E. coli and non- E. coli coliform strains. A 17
total of 50 E. coli isolates and 50 non- E. coli coliform isolates were cultured in ISO Buffered 18
peptone water (BPW), prepared per ISO 6579 (6), at 37 ± 0.5°C for 24 ± 2 h. 19
20
The purpose of exclusivity testing was to determine the ability of the 3M Petrifilm Rapid E.
coli /Coliform Count Plate method to discriminate target organisms from non-target organisms. 21