Microsoft Word - BYLAWS 9_26_10
OMB MEETING MATERIALS
201
RSD i
(Intermediate Precision)
Discussion AOAC itself does not have any guidelines for intermediate precision. ISO 5273- 3 section 0.6 describes four “likely conditions” for intermediate precision: time, operators, equipment, and calibration. The imprecision contained in the ISO definition with its “likely conditions” severely reduces the utility of the resulting data. RSDi, as defined by ISO, is somewhere between RSDr and RSDR, but is not clear where it is between RSDr and RSDR. Hence, there is little more one can say than that the RSDi should be less than RSDR. Since the term is so imprecise it appears there is little to be gained by including RSDi in SMPRs.
08/09/2018