SPSFAM EXPERT REVIEW PANEL
AOAC SPDS ERP REVIEW: MAIN FORM
Submission Date
2017-09-14 22:42:09
Name
Jing Tan
jtan@abbott.com
Organization
Abbott Nutrition R&D
Title of Method
Immunoaffinity purification and HPLC fluorescent detection
AOAC Candidate Method Number (e.g. ALN-01)
BPA-02
Applicable SMPR
BPA_SMPR_V6
Summary:
In this method, immunoaffinity column (IAC) was used for sample cleanup and concentrating, and BPA was determined by HPLC-FLD. It is of high specificity. The method is relatively easy to implement in term of sample preparation and instrument operation and maintenance. Authors have demonstrated most of the method performance according to the definitions in SMPR. But additional information is needed on system suitability and analytical quality control.
1. Does the applicability of the method support the applicability of the SMPR? If not, please explain what is missing. 2. Does the analytical technique(s) used in the method meet the SMPR? If not, please specify how it differs from what is stated in the SMPR. 3. Are the definitions specified in the SMPR used and applied appropriately in the method? If no, please indicate how the terms are used. 4. Does the method, as written, contain all appropriate precautions and warnings related to the method's reagents, components, instrumentation, or method steps that may be hazardous? If no, please suggest wording or option(s).
Yes
Yes
Yes
There is no precaution or warnings written in the method.