2018 March 16 SPDS
51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68 69 70 71 72 73
Repeatability .— Variation arising when all efforts are made to keep conditions constant by using the same instrument and operator and repeating during a short time period.
Expressed as the repeatability standard deviation (SD r
); or % repeatability relative standard
deviation (%RSD r ).
Reproducibility .— The standard deviation or relative standard deviation calculated from among-laboratory data. Expressed as the reproducibility relative standard deviation (SD R );
or % reproducibility relative standard deviation (% RSD R ).
Recovery .— The fraction or percentage of spiked analyte that is recovered when the test
sample is analyzed using the entire method.
Trans Resveratrol.- IUPAC name: 5-[(E)-2-(4-hydroxyphenyl)ethenyl]benzene-1,3-diol. CAS
number: 501-36-0. See figure 1 for chemical structure.
Analytical range – Includes all steps of the analytical procedure including sample
preparation and further dilutions.
5. Method Performance Requirements :
Table 2: Analytical Range & LOQ Based on Matrix
Parameter Analytical range (% w/w)
0.05 - 100
Limit of Quantitation (% w/w)
0.01
74 75 76 77
Range may be narrower depending on the analytical matrix.
Table 3: Method Performance Requirements as a Function of Range
Parameter
Acceptance Criteria
1-50%
50-100%
< 1%
98 - 102
97 - 103
% Recovery
85 - 115
5.3
2
% RSD r
≤ 7.5
≤ 8
3
% RSD R
≤ 10
78 79 80 81 82 83 84
6. System suitability tests and/or analytical quality control:
Suitable methods will include blank check samples, and check standards at the lowest point and midrange point of the analytical range. A control sample must be included. The
method must be specific for trans resveratrol.
Made with FlippingBook - professional solution for displaying marketing and sales documents online