AOAC SPIFAN Nutrients ERP-Final Review (Jan. 2019)

FOR EXPERT REVIEW PANEL USE ONLY NOT FOR DISTRIBUTION

7 samples were in the upper SMPR range of 20 – 200 mg/100g  RSD r ranged from 1.3 – 3.0% • The requirement of RSD r < 5% was thus met;  RSD R ranged from 5.2 – 11.4%; • The requirement of RSD R

< 10% was met for 6 of 7 samples: o 1 sample in the upper concentration range exhibited an RSD R

value of

11.4%

• System suitability:

Recommendation: Recommend for approval as Final Action for Total Choline.