2016_Imaging for Physicists
System measurements
Paper
Year
Phantom size
System
Error
Vermandel
2014 (Head) 24 cm Various 3.0T & 1.5T Mean 1 mm uncorrected, 0.5 mm corrected
Glide-Hurst
2014 2500 points,
1.0T Panorama
24% > 3mm at 150-200 mm radius
40 x 40 x 40 cm
Balter
2014 4689 points,
3.0T Skyra
<1 mm at 17 cm radius (60 mm z) 250 mm radial, 200 mm z 29-87% of phantom <2mm
46 x 35 x17 cm
Walker
2015 5830 points,
3.0T Skyra
50 x 38 x 51 cm
Torfeh
2015 357 rods,
1.5T GE 450w
95% <1mm at 200 mm radial
40 x 40 x 19 cm
Made with FlippingBook