Steps to initiate radiation
Sequence of steps to initiate irradiation includes a test of beam monitoring chambers, but … … the information about missing power supply can not pass through faulty diode … … interlock is not informed that monitoring chambers are missing … and gives green light to the next step in the sequence towards irradiation
Signal is transferred through the diode
A
Signal of failure of the beam monitoring system
Interlock activated prevents irradiation
Diode fails (open circuit) signal not transferred
B
X
Signal of failure of the beam monitoring system
Interlock not activated: cannot prevent irradiation
Function of diode D 29
A: Diode working properly B: Diode disabled (open circuit)