2016 Risk Course Book

37

Looking around

 Copenhagen – QC showed 5% deviation in output - was adjusted immedeatly o Linac OK but incorrect calibration factor for ion chamber – detected after several weeks even if in-vivo dosimetry was in placed (however, lack of comprehensive analysis) o No second physicist checked QC  Ottawa – Recommissioning of unit after move – missed back scatter factors o No second physicist o Detected when annual QC was done  Touluse/Ohio – Commissioning of SRT with unsuitable ion chamber

o No second physicist o Detected by company