ESTRO course Dose Modelling and Verification 2017
11
Inappropriate commissioning
Reported 2007 at Hôpital de Rangueil in Toulouse, France
In April 2006, the physicist in the clinic commissioned the new BrainLAB Novalis stereotactic unit o Possible to use small fields (6x6mm) o “…an ionisation chamber of inappropriate dimensions…” for calibrating the smallest microbeams was used ( Farmer chamber was used) o The incorrect data was entered into the TPS o 145 patients affected
Warsaw 2017