ESTRO course Dose Modelling and Verification 2017
38
Positioning rounded collimator edges
Important for dose calculations in small fields and IMRT.
Focal point
A-C
B-C is nearly constant (approx. 0.3 mm) for a Varian MLC.
A-B
A: Projected tip
B: Tangent (light field)
C: Half Value Transmission
Boyer and Li [6]
A,BC
CB A
T Knöös
Warsaw 2017
Made with FlippingBook - professional solution for displaying marketing and sales documents online