41
Leaf design in the width direction
The leaves are thicker at the base in order to follow the divergence of the beam.
No backup collimators in place!
Inter-leaf leakage is minimized through “tongues” and “grooves”.
6 MV photons
Siemens 160 MLC & Elekta Agility
Siemens 160 MLC
Huq et al [5]
T Knöös