ESTRO course Dose Modelling and Verification 2017
15
Lessons learned from Epinal
Potential errors o Wrong use of TPS due to lack of training + unsafe screen display #1 o Dose due to verification imaging (MV portal) not taken into account #2 o Calculation error due to in-house software, not tested, not qualified #3 o Sole physicist Prevention o Time and organisation for continuous training o Team of physicists (at least 2) o QA for software o Software with safe human-computer interaction o In vivo dosimetry and second independent calculation
Warsaw 2017
Made with FlippingBook - professional solution for displaying marketing and sales documents online