Microsoft Word - 3M Petrifilm REC Plate OMA Manuscript 09121

ERP Methods Book

67

Evaluation of the 3M™ Petrifilm™ Rapid E. coli /Coliform Count Plate Method  1

2

3

AOAC® Performance Tested Method SM 051801 

4

5

Abstract 

6 The 3M™ Petrifilm™ Rapid E. coli/ Coliform (REC) Count Plate is a selective and differential  7

sample‐ready‐culture‐medium system used for the rapid enumeration of Escherichia coli and  8

coliforms from a broad range of foods and select environmental surfaces. The method was  9

evaluated for AOAC® Performance Tested Method SM  certification in accordance with AOAC  10

guidelines. Results from the inclusivity and exclusivity evaluation showed that the test method  11

can distinguish E. coli  from all non‐ E. coli organisms. For non‐ E. coli  coliform strains, the 3M  12

Petrifilm Rapid E. coli /Coliform Count Plate can differentiate coliforms from non‐coliform  13

strains. In the matrix study, the performance of the 3M Petrifilm Rapid E. coli /Coliform Count  14

Plate was compared to the U.S. Food and Drug Administration reference method for 31 foods  15

and two environmental surfaces, and to the ISO reference methods for 12 of the 31 foods and  16

both surfaces. Mean differences calculated between the 3M Petrifilm Rapid E. coli /Coliform  17

Count Plate and each reference method were <0.5 log 10  for each level of each matrix. In  18

robustness testing, small deviations to sample volume and incubation time did not affect test  19

results. The 3M Petrifilm Rapid E. coli /Coliform Count Plate demonstrated reliability as a rapid  20

and accurate alternative to the reference methods for the enumeration of E .  coli and coliforms  21

22

for all foods and environmental surfaces tested in this study. 