Calculation:
For each standard solution, the response factor of each compound is calculated as follows:
R = A x M° M x A°
R
=
response factor of the compound weight of the compound in mg/50 cm 3
M = A = M° = A° =
peak area of the compound weight of IS in mg/50 cm 3
peak area IS
For each compound a calibration curve of the type, y=ax, will be obtained. With this curve all the concentrations of the compounds present in the sample can be calculated. (r2= min 0.995) Formula:
C = M° x A x 100 a A° W
C = concentration of the compound in mg/100 g sample A = peak area of the compound M° = weight of IS in mg/50 cm 3 A° = peak area IS W = weight of sample in g/50 cm 3
The reintegration and calculations are performed with the software ATLAS or similar.