Henry Schein Dental. ACE CATALOGO INFINITY TRICAM 2019
i n f i n i t y TR I - CAM imp l an t s y s t em
21
TRI-CAM Drilling Sequences *
A
B
C
D
E
TRI-CAM Surgical Drilling Protocols
Ø 1.6mm Round Drill Drill pilot hole
Ø 2.0mm Regular Drill Drill to implant length
Paralleling Pin
3.5mm Tapered Drill Match to implant length
4.3mm Tapered Drill Match to implant length
5.0mm Tapered Drill Match to implant length
Paralleling Pin
Platform Dense Bone Drill Match implant diameter
Platform Bone Tap
Delivery Driver
Match implant diameter
Match implant diameter
Match implant diameter
Match platform color
800- 1200 RPM
800- 1200 RPM
800- 1200 RPM
800- 1200 RPM
800- 1200 RPM
800- 1200 RPM
20 RPM 45 Ncm
20 RPM 45 Ncm
SPEED
–
–
Optional Dense Bone
Optional Dense Bone
Drill to the top of the TRI-CAM tapered drill cutting edge, as shown
Ø 5.0 mm (ps) with 4.3mm platform
Optional Dense Bone
Optional Dense Bone
Ø 5.0 mm
* Caution: The tapered surgical drills are 1.5mm longer in length than what is described. * Note: Instruments shown are drilling for a 10mm implant. Adjust accordingly for other implant depths.
Drill Laser Markings
Profile Drill Markings
Bone Tap Markings
Drill to the top marking on the Crestal Bone Drill.
For implant lengths of 10, 11.5, 13, and 16mm drill to top marking. For 8mm implants drill to lower marking.
16mm 13mm 11.5mm 10mm
8mm 6mm
Made with FlippingBook Annual report maker