IGRT 2016 Madrid
2D vs 3D setup
2D vs 3D setup correction
Phantom test: identical Detected error in CBCT larger in all directions CBCT measurements better reproducable
Planar imaging underestimates the setup error, especially if you do it manually Deformation might contribute to this (Li et al)
Fuller et al. 2009
How flexible is the neck ?
How flexible is the neck ?
• Purple: reference CT scan
• PTV conflicts with OAR
• You cannot align all at the same time: Compromise!
• Green: daily
localization Cone Beam CT scan
• Need for choices in alignment!
How flexible is the neck ?
How flexible is the neck ?
31 patients 8 CBCT scans per patient: 249 scans evaluated 9 ROIs per patient Retrospectively: Larynx assumed as PTV
Count occurrence of ROI within certain margin
ROI:
• PTV(larynx)
• OAR (C3-C5)
Margin 0..10 mm (1 mm)
Vertebrae assumed as OAR