CS8000 C 2013 ENG

Technical specifications

Radiological exam options

Pan: Full panoramic, segmented panoramic, maxillary sinus, LA TMJ x 2, LA TMJ x 4 Ceph: Lateral, frontal AP or PA, oblique, submento-vertex, carpus 2 patient types (adult and child) 3 patients sizes (small, medium, large)

Tube voltage

60 - 90 kV

Tube current

2 - 15 mA (max)

Frequency

140 kHz (max)

Tube focal spot

0.5 mm (IEC 60336)

Exposure mode

Total filtration

> 2.5 mm eq. AI

Sensor technology

CCD and optical fiber

Exposure time

Pan: 4 to 14 seconds Ceph: 0.1 to 3.2 seconds 230 - 240 V - 50/60 Hz 100/110/130 V - 50/60 Hz

Sensor matrix

Pan: 64 x 1348 pixels Ceph: 1360 x 1840 pixels

Input voltage (AC)

Image field

Pan: 6.1 x 129.4 mm Ceph: 195 x 263 mm

Unit dimensions

2250 (L) x 1261 (D) x 2315 (H) mm / 88.6 (L) x 49.6 (D) x 91.1 (H) in 2350 (L) x 1950 (D) x 2400 (H) mm / 92.5 (L) x 76.8 (D) x 94.5 (H) in

Gray scale

16384 - 14 bits

Magnification

Pan: 1.27 (± 10%) Ceph: 1.14

Minimum required space

2250mm (88.59")

1261mm

(49.63")

(87.25'')

2216mm

MAXI1687mm (66.41") MINI997mm (39.25")

454mm (17.88'')

1796mm (70.71'')