Assay Temperature showed no significant difference by t-test or paired log-T test confidence levels

>0.5. A shorter assay time did not show a significant difference by t-test and there is no significant 553

difference between the shorter 22 and longer 50 hour incubation times. Pipet volume did show a 554

significant difference by t-test as would be expected with a low bias of the 900µL compared to the 1100 555

µL dispense; however the low bias using the mean log difference and confidence limits >0.5 log 556

557

specification is not considered significantly different.

558

The effect of moisture loss from an exposed unsealed test strip and the effect of moisture loss on a

test exposed for 15 minutes in a laminar flow hood were determined. In control experiments with 559

sealed strips, there is less than a 1% loss of weight, while there was a 10-15% weight loss after 15 560

minutes open air exposure that would simulate an environmental sample of a food plant. Moisture loss 561

studies were performed with 3 lots of EB tests in 3 sets (n=10 each) with Butterfields buffer fortified 562

with either Hafnia alvei ATCC 51815 , Pantoea agglomerans ATCC 27155 or Enterobacter aerogenes 563

ATCC 13048 . Plates had diluted samples added and were sealed (control) and left exposed in 32°C 564

incubator for 15 minutes to achieve volume loss of 10- 15%. Average and standard deviations were 565

calculated and reported in Table 12. Overall there were not a significant differences (±2SD) in the 566

bacterial recovery on the control compared to the air exposed plates on any of the 3 lots of tests 567

568

evaluated.

569

Conclusion

570

571 Pre-collaborative study demonstrate that Peel Plate for Enterobacteriaceae Bacteria (EB) method ERP Use Only incubated at 37 ± 1°C for 24 up to 48 h, without a confirmatory step, selectively detect EB and exclude 573 572

non-EB. Matrix studies of heat processed milk and dairy products, infant formula and cereals, and 574

environmental surfaces, and chicken rinse samples, were not significantly different from the standard 575

24

Made with FlippingBook Learn more on our blog