DeterminationofNa,Mg, P, K, Ca, Cr,Mn, Fe, Cu, Zn, Se, andMoby ICP-MS
c)
After
initiating theplasma, allow the instrument towarmupwhileaspiratinga10%nitricacidblank
solution for at least 15minutes.
d)
After thewarm-upof 15minutes, Select
Instrument>>Tune
in the
ICP-MSTop
window. The
ICP-MSTuning-
Sensitivity
window for sensitivityappears. Select the
NoGasMode
from thedropdownboxof the
TuneFile
window, as shownbelow.
MTE-02
FORWORKINGGROUP/ERPUSEONLY
DONOTDISTRIBUTE