AOAC SPIFAN Stakeholder Panel Meeting Book (March 12, 2020)
Working group activities
Discussed during the calls (major topics)
• Analytes?
2’-FL , 3-FL, LNnT , LNT, 6’-SL, 3’-SL, DFL, (LNFP-I)
• SMPR for standards?
not in the scope of SPIFAN and current SMPRs
• Analytical range per HMO?
based on data for breast milk and IF
• Selectivity / interferences?
possible interferences discussed, incl. probiotics
• Reference materials?
not currently available
Specific Performance Claims: 2’-FL Table 1: Method Performance Requirements a Analytical range 5 – 500 mg/100g Limit of Quantitation (LOQ) 4 mg/100g
85-110% (5-20 mg/100g ) 90-110% (>20 mg/100g)
Recovery
Repeatability (% RSD r ) Reproducibility (% RSD R ) a Concentrations apply to: i) “ready-to-feed’ liquids “as is”; ii) reconstituted powders (25 g into 200 g of water); and iii) liquid concentrates diluted 1:1 by weight.
Made with FlippingBook Annual report