ESTRO Brachytherapy for Prostate Cancer 2018
Needles Check
• Verification of loaded brachytherapy needles.
• Place a film on top of the needles. The radiation from the loaded needles exposes an image in the film. • The film will verify correct loading of seeds and spacers within each needle, or indicate any discrepancies or missing seeds.