Dose Course 2018_Flipping book

Detector

Appropiate for

Be careful

❑ For x-ray depth dose measurements

Check polarity effect for both photon and electron beams Always measure with the same nominal V For micro-chambers the stem signal can be an issue Correct by Sw,air for electron depth dose measurements. Use small volume detectors for scanning

Ionization chambers

Plane-parallel chamber for build-up region

❑ Output factors large fields

❑ Output factors for small fields (micro-chambers)

Effective point of measurement

Check stability temperature during measurements

❑ Depth dose measurements and profiles

Energy dependence

Diodes

Sensitivity variation with accumulated dose

Shielded detector for x-rays

Non-shielded for electron beams

❑ May present sensitivity variation with dose rate

Output factors

Check stability temperature during measurements

Shielded detector standard/large fields

Non-shielded for small fields