Small fields: relative dosimetry - field size factor, S cp
Minimise energy/field size dependence
An approximation to account for the influence of spectral changes between the 10cm 10cm and a smaller field (e.g. 4cm 4cm) on detector response would be to cross-calibrate the small detector against a medium size detector in an intermediate field (smaller than the reference field of 10cm × 10cm);
This is referred to as ‘ daisy-chaining’ by Dietrich & Sherouse MedPhys 38(7), 2011
( ) ( ) M A M A M IC
( ) ( ) ref int
AM A
( )
diode
AS
cp
diode
IC
int