Dose Course 2018_Flipping book
5.2
The thermometer has a traceable calibration.
However this only says that the measured value is correct within an interval of ± 0.2 ° C. What is the underlying distribution??
All one can do is to suppose that there is a symmetric lower and upper bound of the
interval { T - D T + D } and that any value between this interval has an equal probability .
T = 293.25 [K]
T -
T +
probability density
292
293
294
295
temperature
Col 1 vs Col 2