AOAC SPIFAN Stakeholder Program Meeting Book (September 10, 2020)
Preliminary Interlaboratory Study
• Highest RSD R observed at lowest chlorate/perchlorate levels • Most laboratories used LC-MS/MS methods
• No obvious biases other than for one laboratory that did not use labeled ISTD • One laboratory used IC-MS – results poor but laboratory attributed this to sample preparation issues and not instrumental analysis • No observed trends based on laboratory experience • Most participants were manufacturers (50 %) or CROs (40 %) - no government laboratories participated • Laboratories expressed a variety of experience and frequency of testing these types of samples
13
Follow-Up Interlaboratory Study
• Follow up study being conducted to gather additional data • Looking for more laboratories to participate • Looking for government laboratories to participate • Additional soy ingredient sample to be included • Free to participate; laboratories must pay for sample shipment
Participant Signup (Sep ‘20)
Sample Distribution (Jan ’21)
Results Due (Feb ’21)
Call for participation open through Oct 1 at https://qa.nist.gov/hamqap
Preliminary Data Evaluation (Mar '21)
Final Report (Fall ‘21)
14
Made with FlippingBook Publishing Software