September 2015 SPADA Meeting
Comments
Line Numbers Comments
Proposed Response
Are these System False Negative and Positive Rates the same as those to be determined in the presence of environmental matrix materials as indicated in the Method Performance Requirements Table (unnumbered)? ow can a ys em a se‐ os ve a e e determined using spiked environmental matrix materials if the spike is the target organism? Seems like it should be False‐ Negative and the other rate without spike should be False‐Positive. Method Performance Table (unnumbered) H S t F l P iti R t b
40‐46
That is correct.
The intent is that spiked and unspiked environmental matrix samples will be used to evaluate false negative and false positive rates respectively. Propose to revise Tavle IV Part 1 to clarify this requirement.
Section 7
Comments
Line Numbers Comments
Proposed Response
Table II.
Additional species suggested.
For discussion.
In the empty cells suggest putting "Investigator to choose strain(s)" if that is the Agree to insert "Investigator intention. may choose strain(s)"
Table III