• Multimode ODS column: •
Scherzo SS-C18 column (100 x 3 mm)* • Purified Porous Silica / Particle Size 3um / Pore Size 13nm / ODS + Anion Exchange + Cation Exchange Ligands • Separation achieved through a manipulation of eluent pH, organic solvent composition, salt concentration, and temperature. • Enables the retention of ionic species with LC-MS compatible mobile phases • Avoids the use of ion pairing agents in mobile phases
• Provides simple and rugged chromatographic retention
*= Imtakt USA, Portland, OR
Proprietaryand confidential — do not distribute © 2015 Abbott Laboratories
11
MFA‐AOAC, 2015
March17,2015
SAMPLE PREPARATION • For powders:
– Weigh 10.0 0.5 g of powder into a tared vessel and record mass. – Add water to a total weight of 70 5 g (powder + water) and record mass. – Stir until homogenous. • For ready-to-feed liquids and reconstituted powders: – Weigh 5.00 0.20 g liquid into a tared 15 mL centrifuge tube and record mass. – Add 40 L of internal standard working solution and vortex. – Add 50 μL of glacial acetic acid and 5.0 mL methanol, vortex. – Transfer aliquot to microfuge tube and centrifuge at 13,000 rpm for 10 min. – Filter an aliquot of the supernatant into an autosampler vial using a 0.2 um PTFE syringe filter.