AOAC OMB Final Action Recommendation (December 2019)-2016.14
2016.14 (October 2019) – FOS-03 METHOD
FOR ERP USE ONLY DO NOT DISTRIBUTE
17.0
0.5
0
80
20
17.1
0.5
100
0
0
27.0
0.5
100
0
0
27.1
0.5
0
80
20
33.0
0.5
0
80
20
276 Table 2016.14E. Example of quadruple waveform for carbohydrate detection 277
Time, s
Voltage, V
Gain region
0.00 0.20 0.40 0.41 0.42 0.43 0.44 0.50
+ 0.10 + 0.10 + 0.10 - 2.00 - 2.00 + 0.60 - 0.10 -0.10
Off On Off Off Off Off Off Off
278
J.
Chromatographic Conditions (Using PA1)
279
280
If using a CarboPac PA20 column (or equivalent), skip this section.
The HPAEC-PAD system is equipped with the CarboPac PA1 guard (2 x 50 mm, 10 µm) and analytical 281
(2 × 250 mm, 10 µm) columns or equivalent, connected in series. The columns are held at 20 °C and the 282
injection volume is 20 µL. NaOH (300 mM) is added post-column (before the PAD detector) at a flow rate 283
of 0.13 mL/min, using a T-piece and a 125-µL reaction coil (or equivalent). Fructose and glucose are 284 separated using the gradient described in Table 2016.14F . Carbohydrates are detected by pulsed 285 amperometry using a gold working electrode and an Ag/AgCl reference electrode, and an appropriate 286