SPDS EXPERT REVIEW PANEL

B. Is there information demonstrating that the method meets the SMPR Method Performance Requirements using the Reference Materials stated in the SMPR? If not, then specify what is missing and how this impacts demonstration of performance of the method. method performs within the SMPR Method Performance REquirements table specifications for all analytes in the SMPR applicability statement? If not, please specify what is missing and whether or not the method's applicability should be modified. A. Based on the supporting information, were there any additional steps in the evaluation of the method that indicated the need for any additional precautionary statements in the method? B. Does the method contain system suitability tests or controls as specified by the SMPR? If not, please indicate if there is a need for such tests or controls and which ones. C. Is there information demonstrating that the method system suitability tests and controls as specified in the SMPR worked appropriately and as expected? If no, please specify. C. Is there information demonstrating that the

No supporting information was submitted.

No supporting information was submitted.

IV. General Submission Package

No supporting information was submitted.

No system suitability test was found in the manuscript. It is suggested that check standard analyses at lowest, middle range of analytical range and a control sample are added in order to meet SMPR guideline for system suitability test.

No supporting information was submitted.