F. Based on the supporting information, what are the cons/weaknesses of the method?

The stability of the standard is a critical aspect in this method. It might be worth to explore the use of temperature controlled sampler chamber, which can keep the temperature low (5-10oC) so a large number of injections can be done in one sample queue.

G. Any general comments about the method?

No

Recommendation for the Method

Do you recommend this method be adopted as a First Action and published in the Official Methods of Analysis of AOAC INTERNATIONAL? Please specify rationale.

Recommended to be adopted as a First Action method on the condition of providing satisfactory analysis result of the reference material.