In order to check the device is working, ensure the green light is on and that no other lights are flashing. Refer to table below if any other lights are on. If all ok, check the alarm on the device by simply pressing the ‘mode’ button. All lights should flash and alarm beep. Once completed, the display should return to just the green light being illuminated. If the lights fail to come on or buzzer fails to sound during test, test again to double check. If still not working, place a call out for an engineer visit. Monitor display regularly and be aware of any change to the status of the device. Refer to below table for what to do in each circumstance.
Good lamp (green) Alarm 1 lamp (red) Alarm 2 lamp (red) Fault lamp (yellow) Meaning
Action
Check power supply - call engineer for device if not resolved
Off
Off
Off
Off
Power Off
Normal operation
On/Blip off
Off
Off
Off
No action
Isolate gas, place call for engineer to check for gas leak Evacuate area, place call for engineer to check for gas leak
Off
Flashing and slow buzzer
Off
Off
CO2 level >1.5%
Off
Flashing
Flashing and Med Buzzer
Off
CO2 level >4.0%
Place call for engineer visit for device Place call for engineer visit for device Place call for engineer visit for device
Off
Off
Off
Flashing and slow buzzer Calibration error
Off
Off
Flashing
Flashing and med buzzer Detector fault
Off
Flashing
Flashing
Flashing and fast buzzer I.R. Source fault
Manager to record the training