Basic Training 4th Annual ICP Conference
Sulfur in steel
3 – Sample Measurement Optical Emission or Mass Spec?
ICP-OES
ICP-MS
1. Detection limits
dilution factor vs. IDLs
2. Matrix (complex) 3. Wavelengths / Masses 4. Standard compatibility
good options
~monoisotopic, interferences
N/A
N/A
Assume 500 ppm S in sample = 1 ppm S in analytical solution (MS dilution >500x (0.2% TDS)
Sulfur in steel
3 – Sample Measurement Optical Emission or Mass Spec?
ICP-OES
ICP-MS
1. Detection limits (Method) 2. Matrix (complex) 3. Wavelengths / Masses 4. Standard compatibility
IDL (50 ppb) * 500x = 25 ppm MDL
IDL (1 ppb) * 500x = 500 ppb MDL
good options
~monoisotopic, interferences
N/A
N/A
Assume 500 ppm S in sample = 1 ppm S in analytical solution (MS dilution >500x (0.2% TDS)
ICP-MS slight advantage (?)
Estimated MDL ≈ IDL * Dilution Factor
Made with FlippingBook flipbook maker