Figure 20: Plotted raw data for NIST 1869 CRM powder (fortified)

Method Precision

The submitted data was statistically analysed using the AOAC protocol for overall mean, intra-laboratory

repeatability (S

r

), repeatability relative standard deviation (RSD

r

), inter-laboratory reproducibility (S

R

),

reproducibility relative standard deviation (RSD

R

), and HorRat (2). Cochran (p = 0.025, 1-tail) and Grubbs

(single and double, p = 0.025, 2-tail) tests were utilized to determine outliers (3,4). The statistical analysis of

results submitted by participants is given in Table 2 with pertinent values for method illustrated in Figures

21–23. Only a single pair of results for vitamin D

2

(samples GBAN911/PTLO721) from Lab 6 were excluded

as Cochran outliers, no other outliers were identified and all other results were used in the generation of

precision values.

Precision as repeatability ranged from 1.9–5.8% RSD

r

and reproducibility values ranged from 6.4–12.7%

RSD

R

. For all vitamin D fortified products, the repeatability and reproducibility are within limits set in the

Vitamin D SMPR (2). Acceptable reproducibility was also demonstrated with better than expected HorRat

R

values for the method ranging from 0.2–0.6 (expected range 0.5–2.0) (4).

0.0

5.0

10.0

15.0

20.0

0 1 2 3 4 5 6 7 8 9 10 11 12 13

Vitamin D

2

(μg hg

-1

)

Lab Number

Sample GBAN911/PTLO721

2016.05 (FEBRUARY 2017) VITD-18