Previous Page  78 / 162 Next Page
Information
Show Menu
Previous Page 78 / 162 Next Page
Page Background

76 | Chapter 4

Speech perception

Speech discrimination scores were obtained for all 19 adult cochlear implants users in the study group

during normal clinical follow-up at predetermined intervals, starting one week after initial fitting. The

data used for analysis in this study were the scores obtained at two years of follow-up. Four patients had

to be excluded from the analysis, one patient since he was deceased (of a natural cause) 1.5 years after

implantation, one prelingually deaf patient, and two patients for whom only scanning data and no selectivity

measures were available. All subjects used the HiRes processing strategy. The standard Dutch speech test of

the Dutch Society of Audiology, consisting of phonetically balanced monosyllabic (CVC) word lists, was

used (Bosman & Smoorenburg, 1995). To improve test reliability, four lists (44 words) per condition were

administered. Although this test is typically scored with phonemes in the Netherlands and Flanders, the

data are shown as word scores, which is a more common reporting method in Anglo-Saxon countries. All

testing was performed in a soundproof room, using a calibrated loudspeaker in frontal position at one metre

distance. Subjects were tested in quiet at speech levels of 65 dB SPL in a CI-only condition.

468

F. B. van der Beek et al.

Speech perception

Speech discrimination scores were obtained for all 19 adult cochlear

implants users in the study group during normal clinical follow-up

at predetermined intervals, starting one week after initial fitting. The

data used for analysis in this study were the scores obtained at two

years of follow-up. Four patients had to be excluded from the analy-

sis, one patient since he was deceased (of a natural cause) 1.5 years

after implantation, one prelingually deaf patient, and two patients for

whom only scanning data and no selectivity measures were avail-

able. All subjects used the HiRes processing strategy. The standard

Dutch speech test of the Dutch Society of Audiology, consisting

of phonetically balanced monosyllabic (CVC) word lists, was used

(Bosman & Smoorenburg, 1995). To improve test reliability, four

lists (44 words) per condition were administered. Although this test

is typically scored with phonemes in the Netherlands and Flanders,

the data are shown as word scores, which is a more common report-

ing method in A gl -Saxon countries. All testing was performed in

a soundproof room, using a calibrated loudspeaker in frontal position

at one metre distance. Subjects were tested in quiet at spe ch level

of 65 dB SPL in a CI-only condition.

Analysis

Signal processing was performed off-line using Matlab. eCAP

amplitudes were automatically detected using Matlab software

(as per Frijns et al, 2002) and plotted against the electrode positions

along the array. Curves that did not show eCAP amplitudes above

0.1 mV were not included in the analysis. This criterion was not

reached in 16% of the responses, mainly in the low current range.

The average of the peak amplitude, for both selectivity and scanning,

was 0.6 mV. The curves were normalized by taking the value at the

electrode contact of interest and dividing all values along the array

by this value. Next, both flanks of the selectivity and scanning curves

were fitted by a 4th order polynomial.

The width was defined as the number of electrode contacts

(spaced 1.1 mm apart) from the stimulated contact to the point at

which the normalized amplitude reduces to 0.6. For the middle con-

tact both the width in the apical (EM-A) as well as the basal direc-

tion (EM-B) were calculated. In cases where the minimum value

did not drop to 0.6 the width was set as the limit of the array in

the apical or basal direction (as per Abbas et al, 2004). In previous

studies both 50% and 75% of the peak mplitu e hav been used

as a measure of the width of the region of excitation (Cohen et al,

2003; Hu hes & Abbas, 2006 ). Fo this study, 60% of the peak

amplitude, determined on the basis of the fitted, normalized curves,

was selected as a trade-off between obtaining as many curves as

possible and being able to measure differences between distinctive

profiles along the array.

Figure 2 shows typical selectivity curves recorded in one sub-

ject. The figure shows the normalized eCAP amplitudes obtained

0 0.2 0.4 0.6 0.8 1 1.2

-1

0

1

2

3

4

5

6

7

8

Time from stimulus end (ms)

Recorded potentials (mV)

0 0.2 0.4 0.6 0.8 1 1.2

-2

-1

0

1

2

3

4

5

6

7

8

Time from stimulus end (ms)

Recorded potentials (mV)

0 0.2 0.4 0.6 0.8 1 1.2

-0.5

0

0.5

1

1.5

2

2.5

3

3.5

4

Time from stimulus end (ms)

Recorded potentials (mV)

A

B

P

M

Stim

D

C

0 0.2 0.4 0.6 0.8 1 1.2

-0.5

0

0.5

1

1.5

2

2.5

3

3.5

4

Time from stimulus end (ms)

Recorded potentials (mV)

Artefact rejection:

Forward masking

Artefact rejection:

Alternating polarity

Scanning

Selectivity

Patient 1

Patient 2

R16

R15

R14

R13

R12

R11

R10

R9

R8

R7

R6

R5

R4

R2

R1

R16

R15

R14

R13

R16

R15

R12

R11

R10

R9

R8

R7

R6

R5

R4

R2

R1

P16

P15

P14

P13

P12

P11

P10

P9

P7

P6

P5

P4

P2

P1

P3

P16

P15

P14

P13

P12

P11

P10

P9

P7

P6

P5

P4

P2

P1

P3

Excited fibres

Recording

Recording

Recording

Refractory fibres

Masker (M)

Probe (P)

Masked Probe (MP)

M + P - MP

Figure 1.

A: For scanning measures a single probe electrode contact is stimulated (Stim) and the response is recorded at all electrodes

individually. Recordings at electrodes further away from the stimulating electrode results in lower eCAP amplitudes (shown below electrode

array). B: Selectivity measures are made by stimulating a masker electrode contact (M) immediately before the probe contact (P). The neural

population stimulated by the masker will then be in a refractory state when the probe is activated, so that the response to the probe will be

reduced in proportion to the amount of overlap. When the measurements of masker alone and probe alone are summed and the measurement

of masker and probe combined is subtracted (M P MP) the eCAP measurement will show the overlap between the cochlear region excited

by M and P. C & D: Raw waveforms of individual patients are shown with detected peaks marked (N1, circles; P1, squares). C: waveforms of

an individual patients’ scanning data using the forward masking and the alternating polarity artefact rejection method. Each waveform labeled

with corresponding recording electrode (R1, R2, R3…). D: waveforms of selectivity data from two individual patients (as always in selectivity

measurements a forward masking technique was used). Each waveform labeled with the corresponding position of the probe (P1, P2, P3…).

Int J Audiol Downloaded from informahealthcare.com by Leids University on 11/27/13

For personal use only.

Figure 1.

A: For scanning measures a single probe electrode contact is stimulated (Stim) and the response is recorded at all electrodes

individually. Recordings at electrodes further away from the stimulating electrode results in lower eCAP amplitudes (shown below

lectrode ar ay). B: Selectivity measures are made by sti ulating a masker electrode co tact (M) immediately b fore the probe contact

(P). The neural population stimulated by the masker will then be in a refractory state when the probe is activated, so t at the response to

the probe will be reduced in proportion to the amount of overlap. When the measurements of masker alone and probe alone are summed

and the measurement of masker and probe combined is subtracted (M + P - MP) the eCAP measurement will show the overlap between

the cochl ar region excited by M and P. C & D: Raw waveforms of individual patients are shown with detected peaks marked (N1,

circles; P1, squares). C: waveforms of an individual patients’ scanning data using the forward masking and the alternating polarity artefact

rejection method. Each waveform labeled with corresponding recording electrode (R1, R2, R3…). D: waveforms of selectivity data from

two individual patients (as always in selectivity measurements a forward masking technique was used). Each waveform labeled with the

corresponding position of the probe (P1, P2, P3…).