Specifications

HI9931

Range

0.00 to 10.00 mS/cm

Resolution

0.01 mS/cm

Accuracy

±2% f.s.

Calibration

through “ZERO CAL” and “SLOPE CAL” trimmers

Set point

from 0 to 10.00 mS/cm

Temperature

Compensation

automatic, 0 to 50°C (32 to 122°F) with

β

= 2%/°C

Recorder Output

selectable at 0-20 mA or 4-20 mA (isolated)

Proportional Control

conductivity from 0.0 to 1.6 mS/cm and time cycle from 0 to 90 seconds

Alarm Contact

terminal can be configured as normally open or normally closed (isolated output

max. 2A, max. 240V, resistive load, 1,000,000 strokes). The alarm is activated if

conductivity exceeds by more than the user selectable interval (0 to 2.0 mS/cm)

from the set point or due to overdosage

Dosing Terminals

relay (115 to 240V, max.2A, 1,000,000 strokes) is activated whenever conductivity

falls below the setpoint

Probe

four-ring potentiometric with built-in temperature sensor and DIN connector

(not included)

Power Supply

±10% 115 VAC or 230 VAC; 50/60 Hz

Environment

-10 to 50°C (14 to 122°F); RH max 95% non-condensing

Case Materials

fiber-reinforced, self-extinguishing ABS

Dimensions

221 x 181 x 86 mm (8.7 x 7.1 x 3.4’’)

Weight

1.6 kg (3.5 lb.)

Ordering

Information

HI9931

is supplied complete with instructions.

Choose your configuration

HI9931-1

115V

HI9931-2

230V

HI9931

Industrial Grade EC

Controller

with Proportional Fertilizer Dosing for

Hydroponics Applications

• Alarm

·

·

Fail Safe Alarm System

• ATC

·

·

Automatic temperature compensation

HI9931 is a wall mounted meter that

measures and controls conductivity in the

0 to 10 mS/cm range. A single set point

allows for proportional dosage of fertilizer

solutions. The proportional settings can

be fine tuned through two conveniently

positioned dials on the front panel. The time

cycle is adjustable from 0 to 90 seconds and

the proportional band from 0 to 1.6 mS/cm.

Calibration and set points have a coarse and

fine tuning trimmers. A pump or electrovalve

can be wired directly to the controller and be

powered through the terminals.

HI9931 also provides for an alarm relay which

is activated when the measurements exceed

the set point by a user selectable margin from

0.5 to 2.5 mS/cm. The alarm also triggers if,

due to a malfunction, the continuous dosing

time exceeds the operator adjustable period

of 1 to 10minutes. The alarmcanbe configured

in either normally closed or open position and

turned off during maintenance. HI9931 also

provides an isolated output signal which is

user selectable between 0-20 or 4-20 mA.

Hanna four-ring conductivity probes ending

in a DIN connector can be quickly attached

to the HI9931. Readings are automatically

compensated for the effects of temperature

in the 0 to 50°C (32 to 122°F) range. For

quick and easy wiring, HI9931 comes with

extractable terminal modules. Several LED’s

show whether the set point or alarm relays

have been activated.

EC solutions begin on page 5.34