A. Principle

Test samples are extracted with acetonitrile using a

homogenizer, and the extracts are cleaned up with Cleanert

TPT (Agela Technologies, Tianjin, China), EnviCarb/PSA

or ECPSACB506 (UCT, Bristol, PA), BE Carbon 500 g/PSA

(Agilent Technologies, Santa Clara, CA), or InerSep GC/PSA

(GL Sciences, Tokyo, Japan), or equivalent cartridges. The

pesticides are eluted with acetonitrile–toluene (3 + 1, v/v),

concentrated, dried, dissolved in the recommended solution,

and analyzed by GC/MS, GC/MS/MS, or LC/MS/MS.

Quantifications is with a matrix-matched standard calibration

curve.

B. Reagents

(a)

Solvents

.—Acetonitrile, toluene, and

n

-hexane (HPLC

grade).

(b) 

Acetonitrile–toluene.—

3 + 1 (v/v).

(c) 

Ultrapure water

.—Obtained from a Milli-RO plus

system together with a Milli-Q system (EMD Millipore Corp.,

Billerica, MA).

(d) 

Anhydrous sodium sulfate

.—Analytically pure. Baked

at 650°C for 4 h and stored in a desiccator.

C. Materials

(a) 

SPE cartridges

.—Cleanert TPT (2000 mg, 12 mL; Agela

Technologies), or EnviCarb/PSA (500 mg/500 mg, 6 mL;

Supelco, Bellefonte, PA), or equivalent.

(b) 

SPE tube adapter

.—For 12 mL SPE tubes (57267), for

6 mL SPE tubes (57020-U; Sigma-Aldrich Shanghai Trading

Co., Ltd, Shanghai, China), or equivalent.

(c) 

Disposable flow control valve liners

.—For Visiprep

TM-DL (57059; Sigma-Aldrich Shanghai Trading Co., Ltd).

(d) 

Pear-shaped flask

.—80 mL (Z680346-1EA; Sigma-

Aldrich Shanghai Trading Co., Ltd), or equivalent.

See

Figure 

2014.09A

.

(e) 

Reservoir

.—30 mL (A82030; Agela), or equivalent.

See

Figure

2014.09A

.

(f) 

Centrifuge tube

.—80 mL.

(g) 

Millipore filter membrane (nylon)

.—13 mm × 0.2 μm.

D. Preparation of Standard Solutions

(a) 

Prep

aration

of stock solutions.—

Accurately weigh

5–10 mg individual pesticide and chemical pollutant standards

(accurate to 0.1 mg) into a 10 mL volumetric flask. Dissolve and

dilute to volume with methanol, toluene, acetone, acetonitrile,

isooctane, etc., depending on each individual compound’s

solubility. All standard stock solutions are stored in the dark at

0–4°C and can be used for 1 year.

(b) 

Preparation of mixed standard solution.—

Depending on

properties and retention times of compounds, all compounds

are divided into a series of groups. The concentration of each

compound is determined by its sensitivity on the instrument

for analysis. Mixed standard solutions are stored in the dark

below 4°C.

(c) 

Working standard mixed solution in matrix.—

Prepare

working standard mixture solution in matrix of pesticide and

chemical pollutants by diluting an appropriate amount of mixed

standard solution with blank extract, which has been taken

through the method with the rest of the samples. Mix thoroughly.

Used for plotting the standard curve. Working standard mixture

solution in matrix must be freshly prepared.

E. Apparatus and Conditions

(a)

GC/MS analysis.—

(

1

) 

System

.—Model 7890A gas

chromatograph connected to a Model 5975C mass selective

detector with electron ionization (EI) source and equipped with

a Model 7683 autosampler and Chemstation data processing

software system (Agilent Technologies), or equivalent.

(

2

) 

Column

.—DB-1701 capillary column (30 m × 0.25 mm ×

0.25 μm; J&W Scientific, Folsom, CA), or equivalent.

(

3

) 

Column temperature

.—40°C hold 1 min, 30°C/min to

130°C, 5°C/min to 250°C, 10°C/min to 300°C, hold 5 min.

(

4

) 

Carrier gas

.—Helium, purity ≥99.999%, flow rate

1.2 mL/min.

(

5

) 

Injection port temperature

.—290°C.

Figure 2014.09A. SPE equipment.