

* Visit
inorganicventures.com/tech/icp-operations/for additional information from this link
The following example is for an application where a submitter has been obtaining minor levels (0.1 to 1.0 %) of Cr in an
alloy containing roughly equal amounts of Fe and Ni. The laboratory where this alloy is analyzed uses a procedure where 0.2
grams of the sample is dissolved in 5 mL of a 1:1 HNO3 / HCl mixture and diluted to 1000 mL with DI water. The analyst is
informed that a limit of detection (LOD = 3SD
0
) of 1 ppm Cr based upon the original sample and the ability to quantify the
Cr to within ±10 % relative at the 10 ppm level is an absolute minimum requirement.
ćF TVCNJUUFS UIFO BTLT UIF BOBMZTU UIF VTVBM RVFTUJPO i* OFFE UIF SFTVMUT UPNPSSPX o DBO ZPV EP JU w ćF BOBMZTU EPFT B RVJDL
calculation and determines that using the most sensitive Cr line and the current procedure, the lowest possible detection limit
is 4 ppm and a more realistic estimation would be ~ 4 times the IDL or ~ 16 ppm. The analyst then pulls up the following
spectra, instrument detection limits, and linear regression data which were obtained on their radial view instrument about
four years ago when installed using pure single element solutions as described above.
The 205.552 nm Cr line was found to be the most sensitive of the 16 Cr lines originally characterized with an IDL of 4.0 ppm
= [ (0.0008 μg/mL Cr IDL) x 1000 ] / 0.2 based upon original sample size and dilution as described above. However, the
spectrum of a 0.1 ppm Cr standard shows significant interference from both Ni and Fe at a concentration of 100 ppm making
the line useless at low ppm Cr levels (see Figures 7.1 and 7.2).
The analyst then begins the relatively simple process of identifying a Cr line with the most sensitivity that is spectrally clean.
Figures 7.3 and 7.4 show the line identified using the same scan data shown for the 205 Cr line. The 267.716 nm Cr line looks
clean at the current dilution factors and has an IDL of 0.0016 μg/mL Cr which increases the detection limit to somewhere
between 8 to 32 ppm.
The good news is that the 267.716 line looks spectrally clean and the possibility of increasing the sample size while lowering
the final volume by a factor of 100 is possible (i.e., 2 grams sample up to 100 mL using 20 mL of 1:1 HCl/ HNO
3
). The
concentrations of the Fe and Ni in the final solution would be ~ 10,000 μg/mL each. This capability was confirmed when
40,000 μg/mL solutions of both Fe and Ni were scanned as shown in Figure 7.5. These spectral data indicate a realistic
detection of << 1 ppm Cr.
Figure 7.1:
Spectra of pure 100 ppm Fe and Ni solutions, 0.1 ppm Cr
and a water blank at the 205.552 nm Cr wavelength
Figure 7.2:
IDL, BEC and regression data for the 205.552 nm Cr line
Figure 7.3:
Spectra of pure 100 ppm Fe and Ni solutions, 0.1 ppm Cr
and a water blank at the 267.716 nm Cr wavelength
Figure 7.4:
IDL, BEC and regression data for the 267.716 nm Cr line