r

0

= 1cm

Dose calculation – AAPM-TG43: The Dose Rate Constant

Λ is expressed in units of cGy.h

-1

.U

-1

and

1U = 1 µGy.m².h

-1

= 1

cGy.cm

².h

-1

The dose rate constant, Λ, is defined as the dose rate to water in water at the reference

point, namely at a distance of r

0

= 1 cm on the transverse axis (θ

0

= 90

0

), per unit air

kerma strength, S

K

:

K

0 0

S

)θ,r(D

Λ

(r

0

,

θ

0

)= (1cm, 90

0

)