Commercialised in vivo detectors-diodes
EDP30 EDPHL-3G P30
QED 1116 Isorad-p 1164 Isorad-3 PTW T60010H
Manufacturer
IBA
IBA
ONCOlog
Medical
Sun Nuclear
Sun Nuclear
Sun Nuclear
PTW
Die - type
p
p
n
p
p
n
p
Die – dimensions-area (mm
2
)
1.76
1.76
6.8
0.64
1.54
1.54
1
Die- thickness (μm)
60
60
100
-
-
-
2.5
Build up cap -shape
hemispherical
hemispherical
cylindrical
hemispherical
cylindrical
cylindrical
hemispherical
Build up cap - material
Tantalum Tantalum Tungsten
Brass
Tungsten
Tungsten
Tungsten
Build up cap- water equivalent
thickness (mm)
14
14
30
30.4
26
26
30
Preirradiation dose- modality
8 kGy at 10
MeV
confidential
25 kGy
10 kGy at 10
MeV
10 kGy at 10
MeV
10 kGy at 10
MeV
0 Gy