1.

Basic dose algorithm validation

TPS model comparison tests

o

Needs to be completed for each configured beam.

o

Each physical wedge is a unique beam (different energy fluence

spectrum)

o

Nonphysical wedges can be considered an extension of the

correspondent open field (is not a different beam)

o

No extra measurements are needed. Confirmation that the dose

calculated in the planning module agrees with those in the

modeling module.