3.

IMRT/VMAT calculation validation

o

Accurate dosimetric commissioning of an IMRT system remains a

challenge

o

Results of IMRT credentialling by IROC shows that only 82% of

institutions passed the credentialling end to end test (5%-7mm)

o

Most of the failures were due to the basic beam modeling

inaccuracies.