Detector

Appropiate for

Be careful

Ionization

chambers

For x-ray depth dose

measurements

Plane-parallel chamber for build-

up region

Output factors large fields

Output factors for small fields

(micro-chambers)

Check polarity effect for both photon

and electron beams

Always measure with the same

nominal V

For micro-chambers the stem signal

can be an issue

Correct by Sw,air for electron depth

dose measurements.

Use small volume detectors for

scanning

Effective point of measurement

Check stability temperature during

measurements

Diodes

Depth dose measurements and

profiles

Shielded detector for x-rays

Non-shielded for electron beams

Output factors

Shielded detector standard/large

fields

Non-shielded for small fields

Energy dependence

Sensitivity variation with accumulated

dose

May present sensitivity variation with

dose rate

Check stability temperature during

measurements