Origin of peripheral dose contributions

Anders A

.

Patient scatter

– Unavoidable

– Decrease from target essentially as:

n

=1 column shaped scatter source

n

=2 “blob” shaped scatter source

Head scatter

Sources are all irradiated parts of the treatment head, main source the flattening filter

Limited by the collimating devices

Machine design dependent (FFF machines have less!)

Leakage, including scatter leakage

Depends on collimator and shield thicknesses and material (density)

Collimator design dependent

Treatment technique dependent

Neutrons

Mainly a high Z phenomena at photon energies above photonuclear threshold values

Avoidable by lower beam energy

Only one factor user controllable on a daily basis, all other once per decade (replacement)!