• The signal from the monitor chamber has a component that originates from particles
which have backscattered from the upper part of the jaws into the chamber.
• The fraction of this signal depends on the aperture defined by the upper jaws and
increases as this decreases.
0
0
( ) 1
b
M M M A b A M
M
0
M
b
Monitor chamber
Dose per MU formalisms:
model-based dose calculations
Fraction of the total of the monitor chamber which is
attributed to particles that have backscattered into the
monitor chamber
o
= 1 +
−1
or