Dose per MU formalisms:
Varian Eclipse v13.6 TPS
Overview of Eclipse TPS (29. Feb 2016)
Multi-source beam model that describes:
• primary photon source
• extra-focal photon source
• electron contamination source
• (hard) wedge scatter source
Dose engines
• Pencil Beam Convolution (PBC v10.0.28)
• Pencil kernel - Anisotropic Analytical Algorithm for photons (AAA
v13.6.23)
• Acuros External Beam for photons (Acuros XB v13.6.23)
• Cone Dose Calculation for photons (CDC v13.6.23)
• Generalised Gaussian Pencil Beam for electrons (GGPB v10.0.28)
• Electron Monte Carlo Algoritm (eMC v13.6.23)
• Proton Convolution Superposition (PCS v13.6.23)
Dose Optimisation algorithms for IMRT/VMAT planning
• Photon Optimiser (PO v13.6.23)
• Dose Volume Optimiser (DVO v13.6.23)
• Plan Geometry Optimiser algorithm (PGO V13.6.23)
• Progressive Resolution Optimiser (PRO v13.6.23)
• Nonlinear Universal proton Optimiser v13.6.23
Fluence Delivery modelling algorithms for IMRT planning
• Algorithms that correct optimal fluence maps to deliverable fluence
maps based on constraints imposed by the MLC (Leaf Motion
Calculator, LMC)
Portal Dose Image Prediction (PDIP v13.6.23)
Intermediate dose calculation using: Multi-Resolution
Dose Calculation Algorithm (MRDC v13.6.23)
Depending on MLC motion (static or
dynamic), the total MU of a beam
are adjusted appropriately