EPID in vivo dosimetry-commercial solutions

Product Name

How they work

Dosimetry Check

(Math Resolutions)

From EP images they calculate the fluence at a

plane before the patient.

This fluence is used instead that calculated by a

beam model to calculate the dose to patient

(dose engine)

EpiGrayTM

(DosiSoft) (IBA)

The EP image is converted to EP dose

(deconvolution by scatter from the portal and

patient), This “primary dose” is backprojected

into the patient (attenuation) and convolved with

a scatter kernel.

iViewDose

(Elekta)

The EP image is converted to EP dose

(deconvolution by scatter from the portal and

patient), This “primary dose” is backprojected

into the patient (attenuation) and convolved with

a scatter kernel.

PerFraction

(Sun Nuclear)

3D dose reconstruction

: Beam model +syntetic

MLC logfiles (from EP images)+ CBCT

2D transmission dose comparison

: Predicted

transmission dose (beam model+ dose

calculation algorithm) compared with the

transmission dose measured by the EPID.